Baoqing Sun,Mahler Michael, Mei Jiang,Jing Li, Nanshan Zhong.[J].南京医科大学学报,2008,28(5):273~278
Comparison between two commercial immunoassays:Dr. Fooke ALLERG-O-LIQ versus Phadia ImmunoCAPR System in detecting allergen-specific IgE and total IgE values
投稿时间:2008-08-13  
DOI:10.7655
中文关键词: 
英文关键词: specific IgE  allergy diagnostics  allergen
基金项目:
作者单位
Baoqing Sun Guangzhou Institute of Respiratory Diseases, the First Affiliated Hospital of Guangzhou Medical College, Guangzhou 510120,Guangdong province, China 
Mahler Michael Dr Fooke Laboratorien GmbH, Neuss, Germany 
Mei Jiang Guangzhou Institute of Respiratory Diseases, the First Affiliated Hospital of Guangzhou Medical College, Guangzhou 510120,Guangdong province, China 
Jing Li Guangzhou Institute of Respiratory Diseases, the First Affiliated Hospital of Guangzhou Medical College, Guangzhou 510120,Guangdong province, China 
Nanshan Zhong Guangzhou Institute of Respiratory Diseases, the First Affiliated Hospital of Guangzhou Medical College, Guangzhou 510120,Guangdong province, China 
摘要点击次数: 2740
全文下载次数: 2839
中文摘要:
      
英文摘要:
      Objective: Beside patient’s history and skin prick testing(SPT) the detection of specific IgE(sIgE) represents an important tool of allergy diagnostics. In recent years different technologies for the detection of sIgE have been developed. The objective of this study is the comparison of the ALLERG-O-LIQ with the ImmunoCAPR System using seven inhalant and four food allergens. Methods: Sera from patients were collected and tested for sIgE to inhalant(d1, d2, d5, i6, e1, e5 and m3) and food allergens(f1, f2, f24, f24) by ALLERG-O-LIQ(Dr. Fooke Laboratorien GmbH) and by ImmunoCAPR System(Phadia). Further, samples were also tested for total IgE in both systems. Results: Prevalence of positive test results varied between 0/20(f24) and 11/20(e5) for ALLERG-O-LIQ and between 3/18(f23) and 11/20(d1/d5) for ImmunoCAPR. The qualitative agreement between both methods was found between 75%(f24) and 100%(d2) depending on the allergen. Overall qualitative agreement for inhalant(n = 140), food(n = 78) and all allergens(n = 218) tested was 92.1%(kappa = 0.84), 83.3%(kappa = 0.58), 89.0%(kappa = 0.77), respectively. Sensitivity, specificity, positive predictive value(PPV), negative predictive value(NPV) and diagnostic efficiency(DE) were found at 88.2%, 95.8%, 95.2%, 89.6%, 92.1%(inhalant allergens), 62.5%, 92.6%, 78.9%, 84.7%, 83.3%(food allergens) and 81.5%, 94.4%, 91.5%, 87.5%, 89.0%(all allergens). Conclusion: Good to excellent qualitative agreement between ALLERG-O-LIQ and ImmunoCAPR for the detection of specific and total IgE could be observed. The degree of agreement depended on the allergen and was higher in the group of inhalant allergens. The ALLERG-O-LIQ System represents a reliable test for the detection of specific and total IgE.