The relationship between image quality and CT dose index in spiral CT for the brain of rabbit
DOI:
CSTR:
Author:
Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    Objective:To explore the feasibility and rationality of low-dose spiral CT scan in the examination for the head of rabbit and provide the basis for the low-dose CT scanning in the head of neonate. Methods:The technical parameters were 120 Kv,3 mm slice thickness and every slice was scanned using tube current of 200,110,70 mAs. Five same regions of interest were measured on every image. The homogeneity and noise level of CT were appraised. The technical parameters were 120 kV,3 mm slice thickness. 200,110,70 mAs were employed in each same slice. CT images were evaluated by four doctors using single-blind method,and 3 degrees including excellent image, good image and bad image. Results:The phantom measurement showed that the lower of CT dose and the lower of homogeneity. The lower of CT dose and the higher of noise leve1. The image quality had no significant diference between five conditions(pitch 1.0,200,110,70 mAs;pitch 1.5,200,110 mAs),(P > 0.05). Conclusion:Low-dose spiral scan for examinating the head of rabbit might not only guarantee image quality but also reduce the radiation during CT scan.

    Reference
    Related
    Cited by
Get Citation

王天乐,龚沈初,张树清,沈云霞,桑军阳.兔颅脑螺旋CT扫描吸收剂量与图像质量的关系[J].南京医科大学学报(自然科学版英文版),2007,(12):1487-1490.