Evaluation of the size of artifacts under 1.5T MRI different sequence parameters caused by titanium alloy implants
CSTR:
Author:
Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    Objective:To compare the variation of titanium implant artifacts in different sequences and optimize scanning parameters for high-quality images. Methods:Titanium water phantom was designed for routine sequences and those with optimized parameters scanning;the variation of artifacts was observed with altering field of view,echo train length,bandwidth in sagT2-FSE sequence. Results:Artifacts were changed under different sequences. The reduction of field of view,the decrease of echo train length and bandwidth reduced the artifacts. Conclusion:Change of the scan sequence or scan parameters could reduce artifacts caused by metal implants,improve image quality,and benefit clinical diagnosis.

    Reference
    Related
    Cited by
Get Citation

王传兵,邹月芬,胡志毅,吕 征,李大鹏,储 彬.钛合金植入物在1.5T场强下不同序列参数伪影大小探讨[J].南京医科大学学报(自然科学版英文版),2013,(11):1600-1603.