Evaluation of the size of artifacts under 1.5T MRI different sequence parameters caused by titanium alloy implants
CSTR:
Author:
Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    Objective:To compare the variation of titanium implant artifacts in different sequences and optimize scanning parameters for high-quality images. Methods:Titanium water phantom was designed for routine sequences and those with optimized parameters scanning;the variation of artifacts was observed with altering field of view,echo train length,bandwidth in sagT2-FSE sequence. Results:Artifacts were changed under different sequences. The reduction of field of view,the decrease of echo train length and bandwidth reduced the artifacts. Conclusion:Change of the scan sequence or scan parameters could reduce artifacts caused by metal implants,improve image quality,and benefit clinical diagnosis.

    Reference
    Related
    Cited by
Get Citation

Wang Chuanbing, Zou Yuefen, Hu Zhiyi, Lv Zheng, Li Dapeng, Chu Bing. Evaluation of the size of artifacts under 1.5T MRI different sequence parameters caused by titanium alloy implants[J].,2013,(11):1600-1603.

Copy
Related Videos

Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
History
  • Received:May 17,2013
  • Revised:
  • Adopted:
  • Online: November 25,2013
  • Published:
Article QR Code