Biomechanical research of combined fixation with multiple locking plate osteosynthesis for hyperextension tibial plateau fractures
Author:
Affiliation:

Department of Orthopaedics, the Affiliated People's Hospital of Jiangsu University, Zhenjiang 212000 ;School of Medicine, Jiangsu University, Zhenjiang 212013 , China

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    目的:评估多种锁定接骨板组合固定过伸型胫骨平台骨折(hyperextension tibial plateau fracture,HTPF)的生物力学效能差异。方法:使用标准化人工仿生胫骨构建HTPF模型,分别装配6种接骨板固定方案[前外侧组(第1组)、前内侧组(第2组)、前外侧+前内侧组(第3组)、前内侧+后内侧组(第4组)、前外侧+后内侧组(第5组)、3块接骨板固定组(第6组)]。9个模型于胫骨平台上表面施加250、500、1 000、1 500 N的纵向载荷,通过三维摄像系统计算不同固定方案时的综合位移。9个模型以0.5°/s的角速度匀速扭转近端骨块,记录内旋10°及外旋10°时不同固定方案所需的扭矩。采用重复测量方差分析评价固定方式与压缩/旋转及其交互作用对内固定的影响。采用配对t检验进行组间两两比较,并采用Bonferroni法校正多重比较的检验水准,计算配对标准化效应量Cohen's d及其95%置信区间。结果:纵向加压试验中,载荷越大综合位移量越大,纵向载荷为1 500 N时,各固定方案组的综合位移排序为:第1组[(5.16±2.39)mm]>第2组[(4.04±1.40)mm]>第3组[(2.56±0.57)mm]>第4组[(1.75±0.71)mm]>第5组[(0.70±0.27)mm]>第6组[(0.34±0.10)mm]。水平扭转试验中,旋转10°时的平均扭矩排序为:第1组[(9.87±2.07)N·m]<第2组[(10.61±1.91)N·m]<第4组[(24.72±4.75)N·m]<第3组[(27.61±3.06)N·m]<第5组[(32.81±6.62)N·m]<第6组[(43.95±8.23)N·m]。双板固定方案中,1 500 N载荷下第3、4组的综合位移值显著高于第5组(P<0.05)。双板固定方案中,第3组与第4组、第3组与第5组间扭矩无显著差异(P>0.05);第5组的扭矩值显著高于第4组(P<0.05)。结论:对于关节面完整的HTPF,锁定接骨板的固定效能随接骨板数量的增加而提升。在本研究模型和加载条件下,双板方案中,前外侧+后内侧固定表现出较小的位移,抗扭转能力在数值上较高,其固定效能可能更具优势。

    Abstract:

    Objective: To evaluate the differences in biomechanical performance of using multiple locking plate osteosynthesis for hyperextension tibial plateau fracture(HTPF). Methods: HTPF models were constructed using standardized artificial biomimetic tibia. Six plate fixation schemas were assembled: anterolateral, anteromedial, anterolateral + anteromedial, anteromedial + posteromedial, anterolateral+posteromedial, and tri-plate schema. Nine models were subjected to longitudinal loads of 250, 500, 1 000, and 1 500 N applied to the superior surface of the tibial plateau. The comprehensive displacement for each fixation schema was calculated using a three-dimensional camera system. Nine models were subjected to a uniform rotation of the proximal bone segment at a constant angular velocity of 0.5°/s. The torque required to achieve 10° of internal rotation and 10° of external rotation was recorded for each fixation schema. A repeated-measures analysis of variance was employed to assess the effects of fixation modality, compression(rotation), and their interactions on internal fixation outcomes. Pairwise comparisons were performed via paired t-tests, with multiple comparison adjustments implemented using the Bonferroni correction. Additionally, the paired standardized effect size (Cohen's d) and its 95% confidence interval were calculated. Results: The axial load tests revealed that the greater the applied pressure, the larger the overall displacement. When the axial load was 1500 N, the comprehensive displacement values of the groups of fixation schemas were as follows: anterolateral (5.16 ± 2.39) mm>anteromedial (4.04 ± 1.40) mm>anterolateral + anteromedial (2.56 ± 0.57) mm>anteromedial + posteromedial(1.75±0.71)mm>anterolateral+posteromedial(0.70±0.27)mm>tri-plate schema(0.34±0.10)mm. The horizontal rotation tests revealed that the average torque required for a ±10° rotation was as follows: anterolateral(9.87±2.07)N·m0.05); by contrast, the torque value of Group 5 was significantly higher than that of Group 4(P<0.05). Conclusion:For HTPF with intact articular surfaces, the fixation efficacy of locking plates increases with the number of plates used. Under the model and loading conditionsof this study, the dual-plate schema combining anterolateral and posteromedial plates exhibited smaller displacement and numerically higher torsional resistance, indicating a potential advantage in fixation performance.