DOI:
作者:
作者单位:

作者简介:

通讯作者:

中图分类号:

基金项目:


An Implanted Closed-loop Chip System for Heart Rate Control: System Design and Effects in Conscious Rats
Author:
Affiliation:

Fund Project:

The study was supported by grant from National Nature Science Found (30670767)

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    Abstract:

    Objective: To evaluate the efficiency of an implanted chip system for the control of heart rate (HR). Methods: The HR was recorded in six conscious Sprague-Dawley (SD) rats. An implanted chip system was designed to regulate the HR by stimulating the right cervical vagus nerve according to the feedback of real time HR. Each rat was subjected to 30-min regulation and 30-min recovery. The change of HR during the regulation period was compared with the control. The ECG was recorded during the experiment for 24 h. Results: The ECG signals were successfully recorded during the experiment. The HR was significantly decreased during the period of regulation compared with control (-79.3 ± 34.5, P < 0.01, n = 6) and then recovered to normal after regulation. Conclusion: The described implanted chip system can regulate the HR to a designated set point.

    参考文献
    相似文献
    引证文献
引用本文

Yuxuan Zhou, Yuan Yuan, Juan Gao, Ling Yang, Feng Zhang, Guoqing Zhu, Xingya Gao.[J].南京医科大学学报(自然科学版),2010,(2):107-114

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:
  • 最后修改日期:
  • 录用日期:
  • 在线发布日期:
  • 出版日期: